Installation Method And Instructions For The Silicone Heater Of Blood Analyzer
Today we will give you a brief introduction to the installation method and instructions for the silicone heater of the blood analyzer.
Installation method
1. Pressure sensitive adhesive can be used to bond flat and smooth workpieces.
2. The use temperature of pressure sensitive adhesive: 150℃ continuous, 230℃ instantaneous. Where the power density does not exceed 0.9W/c㎡.
3. The silicone rubber heater coated with pressure-sensitive adhesive should be used within half a year after leaving the factory, otherwise it will affect the performance of the glue.
4. Small workpieces can be prefabricated and converted in the factory to ensure the service life of the heater.
Instructions for use
1. When using this type of electric heating device, it must be noted that its continuous working temperature should be less than 240℃, and the instantaneous temperature should not exceed 300℃.
2. The silica gel electric heating device can work and be pressed, that is, the auxiliary pressure plate is used to make it close to the heated surface. The heat conduction is good, and the current density can reach 3W/c㎡ when the temperature in the working area does not exceed 240℃.
3. The allowable working temperature is less than 150°C under the condition of stick-on installation.
4. In the case of dry burning in the air, due to the temperature resistance of the material, its power density should be less than 1 W/c㎡; in non-continuous working conditions, the power density can reach 1.4 W/c㎡.
5. The working voltage selection is based on the principle of high power-high voltage, low power-low voltage, and special needs can be listed.